Estudio experimental para establecer los valores de torque de inserción en implante dental

The  main  purpose  of  this  research  was  the development of a study to determine the values of insertion torque recommended to reduce the likelihood of generating pathological overload of mandible molar region due to insertion torque applied to adjust the dental implant. Ti6Al4V dental implants...

Descripción completa

Autor Principal: López-Gualdrón, Clara Isabel; Ph. D.(c). Grupo de Investigación INTERFAZ. Universidad Industrial de Santander. Bucaramanga
Otros Autores: Galeano-Arrieta, Carlos Fernando; Cirujano oral y maxilofacial Universidad Industrial de Santander Bucaramanga, Pinillos, Julio Cesar; M.Sc. Grupo de Investigación INTERFAZ. Universidad Industrial de Santander. Bucaramanga, Moreno, Juan Carlos; M.Sc. Grupo de Investigación INTERFAZ. Universidad Industrial de Santander. Bucaramanga
Formato: info:eu-repo/semantics/article
Idioma: spa
Publicado: Universidad Santo Tomás. Seccional Bucaramanga 2014
Materias:
Acceso en línea: http://revistas.ustabuca.edu.co/index.php/ITECKNE/article/view/514
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
Sumario: The  main  purpose  of  this  research  was  the development of a study to determine the values of insertion torque recommended to reduce the likelihood of generating pathological overload of mandible molar region due to insertion torque applied to adjust the dental implant. Ti6Al4V dental implants were manufactured to be inserted  into  bone  tissue  specimens  of  human  cadaver jaw and similar densities of bone substitute specimens Sawbone ® normalized like bone density D2 type. Then were  measured  the  insertion  torque  data  recording  to  adjust the implant at samples. Data from the test were used in simulations based on the finite element method. Through the elastic strain criteria, it was known the strain values less than 0,4% of the elastic strain at the bone implant interface. Based on outputs case study, the values of insertion torque to be applied to insert implants surgery are recommended.